Change in PSNR According to Kvp, Mas and SID in CR System and DR System

  • Authors

    • Kyoungho Choi
    • Jae Young Jung
    • Hyang Ji Jung
    • Eun Jeong Heo
    • Jin Kyu Ko
    • Jung Keun Cho
    2018-06-08
    https://doi.org/10.14419/ijet.v7i2.33.17928
  • CR system, DR system, PSNR, kVp
  • F/S System could not obtain the image valuable for diagnostics unless it was taken under the precise exposure condition according to the patient's somatotype but CR and DR systems can obtain good quality of image by the post-processing after obtain the image under sufficient exposure condition. However, it may expose the patient more than required as it can obtain the image not under the best condition. In this study, the results of assessing the chest images obtained by the changes in mAs, kVp and SID quantitatively represented high PSNR of SID 180 cm, 120 kVp and 3.2 mAs in common in the CR and DR systems. In DR system, the quality of image shows clear differences by the exposure condition depending on the presence of AEC. Particularly, since DR system has marginal condition that represents the difference if the quality of image is good or bad, the study on the optimum exposure condition should be made by equipment.

     

  • References

    1. [1] Dongyoung, K., Jihae, L., Myungsoo, K., Boram, H., Cheonhee, L., Soyeong, K., Sohyun, A., Rena, L., 2011. Setting up a CR Based Filmless Environment for the Radiation Oncology: Korean journal of medical physics, 22(3): 155-162

      [2] Gwangho, J., Yeonghan, K., Busun, K., 2008. A Study on the Exposure Parameter and Patient Dose for Digital Radiography System in Dae Goo: Journal of radiological science and technology, 31(2): 177-182.

      [3] Hore, A., Ziou, D., 2010. Image Quality Metrics: PSNR vs. SSIM: International Conference on Pattern Recognition, (4): 2366-2369

      [4] Hyeon, A., Changsoo, K., Junghoon, K., 2015. Evaluation of Image Quality for Radiographic Positioning using IEC Radiation Quality in the Digital Radiography System: Journal of Contents, 15(7): 289-299.

      [5] Hyunjong, K., Jungwon, K., Soohyung, L., Jooyeon, K., Mooyoung, K., 2015. Radiation Exposure through Private Health Examinations in Korea: Korean Journal of Health Promotion, 15(3): 136-140.

      [6] Jaein, K., Yangseop, L., Dongsoo, J., Mincheol, J., Seungho, B., Kwanseop, L., Dongyun, H., 2011. A Study on the Proper Chest Exposure Conditions of Mobile Digital X-ray Unit by Exposure Index: Journal of the Korean society for digital imaging in medicine, 13(3): 139-144.

      [7] Jincheol, P., Sanghoon, L., 2009. Structural Similarity Based Video Quality Metric using Human Visual System: Journal of broadcast engineering, 14(1): 36-43.

      [8] Jungu, C., 2008. Analysis of Factors that Affect Dose Increase and Image Deterioration in Digital Radiography: Dankuk University

      [9] Keonghsung, L., 1997. A Review of Technical Factors and Evaluation Methods to Improve the Image Quality of Chest Radiography: Journal of Cheju Hanna University, 21: 309-331.

      [10] Kwangjae, L., Mingi, K., Jongwoong, L., Hocheol, K., 2013. Research for The Environmental Optimization of Dose and Image quality in Digital Radiography: Journal of the Institute of Electronics Engineers of Korea, 50(2): 203-209.

      [11] Minsun, Y., Jaeseung, L., Inchul, I., 2012. Evaluation of Approximate Exposure to Low-dose Ionizing Radiation from Medical Images using a Computed Radiography (CR) System: Journal of the Korean Society of Radiology, 6(6): 455-464.

      [12] Sangheok, S., Youngkeun, S., Jaebong, K., 2007. The Comparative Analysis of Exposure Conditions between F/S and C/R System for an Ideal Image in Simple Abdomen: Journal of the Korean society of medical digital imaging technology, 9(1): 37-43.

      [13] Sangkwang, L., Wonyeong, Y., Yeongho, S., 2012. Trends in Visual Quality Assessment: Electronics and Telecommunications Trends, 27(3): 383-89.

      [14] Whal, L., 2011. Current status of medical radiation exposure and regulation efforts: The Journal of the Korean Medical Association, 54(12): 1248-52.

  • Downloads

  • How to Cite

    Choi, K., Young Jung, J., Ji Jung, H., Jeong Heo, E., Kyu Ko, J., & Keun Cho, J. (2018). Change in PSNR According to Kvp, Mas and SID in CR System and DR System. International Journal of Engineering & Technology, 7(2.33), 1148-1151. https://doi.org/10.14419/ijet.v7i2.33.17928